Explain what is meant by spectral, chemical, ionization, andisobaric interference.

Short Answer

Expert verified

When analyte signals collide with signals from other elements in the sample, then the spectral interference happens. When a component of the sample lowers the amount of analyte atomization, then the chemical interference happens. The loss of analyte atoms due to ionisation is referred to as ionisation interference.Isobaric interference is caused by the presence of equivalent mass isotopes of multiple elements.

Step by step solution

01

Define interference.

A substance other than the tested material that can be analysed by the chosen analytical method or can prevent the assayed material from being measured is referred to as an interference. Interferences lead to inaccurate analytical results.

02

Define spectral and chemical interference.

Spectral interference occurs when analyte signals collide with signals from other elements or molecules in the sample, as well as signals from the flame or furnace.

Chemical interference occurs when a component of the sample, through a chemical reaction, lowers the amount of analyte atomization.

03

Define ionization and isobaric interference.

The loss of analyte atoms due to ionisation is referred to as ionisation interference. The ionisation energy of Ba , for example, is smaller than the dissociation energy of BaO .

Isobaric interference is caused by the presence of equivalent mass isotopes of multiple elements in the sample solution.

Unlock Step-by-Step Solutions & Ace Your Exams!

  • Full Textbook Solutions

    Get detailed explanations and key concepts

  • Unlimited Al creation

    Al flashcards, explanations, exams and more...

  • Ads-free access

    To over 500 millions flashcards

  • Money-back guarantee

    We refund you if you fail your exam.

Over 30 million students worldwide already upgrade their learning with Vaia!

One App. One Place for Learning.

All the tools & learning materials you need for study success - in one app.

Get started for free

Most popular questions from this chapter

(This is a long exercise suitable for group work.) Relative intenities for the molecular ion region of several compounds are listed in arts (a)-(d) and shown in the figure. Suggest a composition for each nolecule and calculate the expected isotopic peak intensities.

a) m/z(intensity):94(999),95(68),96(3)

b)m/z(intensity):156(566),157(46),158(520),159(35)

c) m/z (intensity):224(791),225(63),226(754),227(60),28(264),229(19),230(29)

d)m/z(intensity):154(122),155(9),156(12)(Hint:Containssulfur.)

31: Which liquid chromatography/mass spectrometry interface, atmospheric pressure chemical ionization or electrospray, requires analyte ions to be in solution prior to the interface? How does the other interface create gaseous ions from neutral species in solution?

Quantitative analysis by selected ion monitoring. Caffeine in beverages and urine can be measured by adding caffeine- D3as an internal standard and using selected ion monitoring to measure each compound by gas chromatography. The figure shows mass chromatograms of caffeine (m/z,194)and caffeine-DD3(m-z.197), which have nearly the same retention time.

Suppose that the following data were obtained for standard mixtures:

(a) Compute the mean response factor in the equation

role="math" localid="1664874599903" AreaofanalytesignalAreaofstandardsingle=F(concentrationofanalyteconcentrationofstandard)

(b) For analysis of a cola beverage, 1.000mL of beverage was treated with 50.0ฮผLof standard solution containing 1.11 g/Lcaffeine-D3 in methanol. The combined solution was passed through a solid-phase extraction cartridge that retains caffeine. Polar solutes were washed off with water. Then the caffeine was washed off the cartridge with an organic solvent and the solvent was evaporated to dryness. The residue was dissolved in50ฮผL of methanol for gas chromatography. Peak areas were 1144 for m/z197 and 1733 for m/z194. Find the concentration of caffeine (mg/L ) in the beverage.

Explain what is meant by spectral, chemical, ionization, and isobaric interference.

Draw a rearrangement like reaction D in figure 22-12 to show how m/z 58 arises from 4-methyl-2-pentanone.